3.5 Display Settings
58
Use the scaling function to convert input
voltage to the physical units of the measure-
ment parameter for display, such to convert
voltage input for display as electrical current.
Converted values can be displayed in fixed-
or floating-point notation.
Settin
g Analog Channel or Pulse Channel (Revolution)
Scaling
Converting Units (Scaling function)
Normal Display
(Off)
Scaling Enabled
A
mV
1 For the [Cond](condition) setting, select the
desired numerical notation.
Setting options:( default setting)
2 Move to the next item to the right, and
select the scaling method.
Setting options:( default setting)
3 Enter the [Unit] name on the comment entry
screen.
See: 3.6, "Entering Comments" (p. 60)
4 Ratio: Enter the conversion ratio and offset.
2-pt: Enter the values at two points before
and after conversion.
Dec Displays converted values using fixed-point
notation
Exp Displays converted values using floating-
point notation
Off* Scaling is disabled.
Ratio* Convert by specifying the ratio of physical
units per volt of input signal (conversion ra-
tio), offset, and the name of the units.
2-pt Convert by specifying the input signal volt-
age at two points, the physical unit values at
those points, and the name of the units.
12 3
Conversion Ratio Setting
Conversion Ratio Setting
2-Point Setting
2-Point Setting
[Cond: Exp]
[Cond: Dec]
4
A
H
A
L
V
L
V
H
[V]
Actual measurement values (input values)
Converted unit values
Conversion Ratio Setting Method
Units (eu)
Offset
Slope (Conversion
ratio: eu/v)
2-Point Setting Method
Units: [A]
Param 1:V
H
value A
H
value
Param 2:V
L
value A
L
value
Example
V
H
:High value
A
H
:Converted
high value
V
L
:Low value
A
L
:Converted
low value
Combining scaling and display positioning (upper/lower limits)
To use both scaling and upper/lower display limits, set the scaling first, then upper/lower limit val-
ues. With scaling enabled, the upper and lower limit setting ranges are constrained as follows:
-9.9999E+19 Setting value-1.0000E-19, Setting value = 0,
+1.0000E-19 Setting value +9.9999E+19