6.4 Loading Data on the Memory HiLogger
104
Binary waveform data or captured screen images can be reloaded into the
Memory HiLogger.
Loading Waveform Data and Screen Images
Select the File Screen.
The contents of the CF card
are displayed.
1
Apply
The control dialog box appears.
Select a file to load
2
(When loading waveforms)
When the amount of data in a file to be loaded exceeds the internal memory capacity, you
can specify a data point (number) at which to start loading. The loadable data size shows
the maximum number of data points that can be loaded. This setting is not needed if the
data to be loaded will fit in internal memory. Select [OK] to load the data.
Select [Load].
Apply
3
Displays the parent
folder.
Displays the child folder.
After selecting a folder by pressing the ENTER key,
select [Change] in the displayed control dialog box.
Waveform data file name
Waveform data trigger time
Number of data points in the file
Channels saved in waveform data to
be loaded (Analog, Pulse, and Alarm)
Maximum number of data points that
can be loaded (to internal memory)
Settable range of offset to first data
point
Times of first and last data points to
be loaded

Table of Contents