3.3 Setting Configuration 1 – Recording Settings
42
Configure the horizontal axis and recording settings on the [Setting] screen.
To save data while measuring, auto save must
be enabled beforehand (p. 46).
Select the data acquisition interval to suit your measurement objectives. Note
that shorter
recording intervals restrict the maximum recording time.
3.3 Setting Configuration 1 –
Recording Settings
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Select from the listed
options.
1
2
Select the [Setting] screen.
Specify the Recording Interval
Select from the [Interval] setting options.
Setting options:
10ms, 20ms, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s,
10s, 20s, 30s, 1min, 2min, 5min, 10min, 20min, 30min,
1h ( default setting)
Some waveform peaks may not be displayable with
certain interval settings.
For example, when the [Interval] is 1 s and the [Time/
DIV] (time base) is 5 s.
5s/div
1s
Peak
Broken wire detection is automatically disabled when
the interval is set to 10, 20 or 50 ms (p. 51).