9.5 Synchronous Measurements with Multiple Instruments
135
3
9
Chapter 9 External Control
Multiple instruments can be synchronized using the external control terminals.
Parallel synchronization
Set one instrument to master, and set the others
to slave.
9.5 Synchronous Measurements with
Multiple Instruments
GND
EXT.TRIG
TRIG OUT
GND
Connection example
EXT.TRIG
TRIG OUT
GND
EXT.TRIG
TRIG OUT
GND
Master
Slave
Slave
Set the master instrument.
1 Press the SET key to open the [System]
screen.
2 Set the [Trig Out] setting to [Start].
Set the slave instrument.
1 Press the SET key to open the [System] screen.
2 Set the [External Trg In] setting to [Trigger].
3 Press the SET key to open the [Trig&Alm]
screen.
4 Set as follows.
Trig Func: On
Timing: Start
Trig Source Start: OR
Ext Trig Start: On
Turn the start trigger setting for all the channels
(analog, pulse) to
Off. Failure to do so may start the
measurement without synchronization.
Press the START/STOP key of the slave instrument to start the measurement (trig-
ger waiting state). In this state, press the START/STOP key of the master instru-
ment to start the measurement. The slave instrument will start the measurement at
the same time.