3.3 Setting Configuration 1 – Recording Settings
43
3
3
Chapter 3 Setting before Measurement
Displayed waveforms can be expanded or compressed
along the horizontal axis. Shorter setting values corre-
spond to greater magnification. This setting
can be
changed both during and after measurement without
affecting acquired data (p. 83).
Available display time base settings depend on the selected recording interval.
Horizontal axis display options are affected when the recording interval is changed
after measuring, b
ecause the options for expanding (or compressing) the displayed
axis depend on the recording interval used when measuring.
Setting the Display Time Base (horizontal axis magnification, set
as needed)
: Selectable, ×: Not selectable, : Not selectable while measuring
Hori-
zontal
axi
s
Re
cording interval
[ms] [s] [min] [h]
10 20 50 100 200 500 1 2 5 10 20 30 1 2 5 10 20 30 1
100ms
× × × × × × × × × × × × × × × ×
200ms
× × × × × × × × × × × × × × ×
500ms
× × × × × × × × × × × × × × ×
1s
× × × × × × × × × × × × ×
2s
× × × × × × × × × × × ×
5s
× × × × × × × × × × × ×
10s
× × × × × × × × × ×
20s
× × × × × × × × ×
30s
× × × × × × × × ×
1min
× × × × × × ×
2min
× × × × × ×
5min
× × × × × ×
10min
× × × ×
20min
× × ×
30min
× × ×
1h
×
2h
5h
10h
12h
1day
Select from the [Time/DIV] setting options.
Setting options:
100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s,
1min, 2min, 5min, 10min, 20min, 30min, 1h, 2h, 5h,
10h, 12h, 1day [/DIV] ( default setting)
When auto save is enabled for waveforms, 100, 200 and
500 ms settings are automatically changed to 1 s when
measurement starts, and the shorter values cannot be se-
lected while measuring.