3.3 Setting Configuration 1 – Recording Settings
44
Specify how you want recording to start and
stop. The following methods are available.
• Record continuously:
Record continuously between starting and stop-
ping measurement.
• Specified length (time span):
Record only for a specified time span (can be set
to auto
matically repeat) (p. 45)
Selecting the Recording Start/stop Method
Continuous
Recording
Repeat Re-
cording (speci-
fied lengths)
Start
Measurement
Stop
Measurement
To record continuously,
Enable continuous recording (set
[Cont] to
[On]). (default setting)
Recording continues until you press the START/STOP
key, or until a “stop trigger” event occurs.
When auto save is enabled for measurement data,
the saving method depends on whether [Deleting] or
[Split Save] (segmented save) is selected (p. 46).
To record for a specified length (time span),
1 Set [Cont] to [Off].
2 Specify the duration of one recording length
(time span).
Setting options:
d (days) 0 to 499, h (hou
rs) 0 to 23,
min (minutes) 0 to 59, s (s
econds) 0 to 59
1
2
Recording proceeds for the specified length. Record-
ing stops early if you press the
START/STOP key be-
fore the recording length has elapsed, or if a “stop
trigger” event occurs.
Change 1’s digit
Change 10’s digit
Maximum recording length
and data samples
Indicates the maximum record-
ing time to internal memory, and
the maximum number of data
samples. These values are limit-
ed by the specified recording
length and the number of mea-
surement channels used.
See: Appendix 6, "Maximum
Recordable Time" (p.
A11)