4.2 Alarm Output
73
3
4
Chapter 4 Specifying Criteria for Measurements (Triggering, Alarm and Timer)
You can enable beep tones and an alarm output signal (for external use) for each
input channel by setting its alarm criteria. You can also view alarm output status on
screens that display numerical values ([Wave+Value], [Value+Cmnt] and [Value]).
External alarm output requires connection to th
e external con
trol terminals.
See Chapter 9, "External Control" (p. 129) for details.
Select alarm input channels
on the [CH] sc
reen, and specify alarm criteria on
the [Trig & Alm] screen.
4.2 Alarm Output
Alarm Criteria
Channel alarm status (CH1 to CH10 and P1 to P4)
Yellow*: alarm criteria are satisfied
Green: alarm criteria are not satisfied
*. When [Hold Alarm] is enabled (set to [On]),
the indicator remains yellow because the criteria
remain satisfied (until measurement is finished).
The [Buzzer] alarm sound can be set to on or
off. When [All display] is set to [On], the de-
tected waveforms for CH1 to 10 and P1 to 4
set to the alarm source will also be displayed.
Alarm output status (ALM)
Red: alarm signal output active
Green: alarm signal output inactive
1 Select the [Channel] setting option to [ALM].
2 Set a check mark [ ] as shown to enable
alarm output.
3 Set the [Hold Alarm] as required for your
alarm output destination device.
Setting options:( default setting)
4 Enable or disable the alarm beeper by the
[Buzzer] setting.
Setting options:( default setting)
Off
Disable the alarm output signal when alarm
criteria
are no longer met.
On Hold the alarm output active state until mea-
surement is finished.
Off
The Memory HiLogger will not sound an au-
dible beep during alarm signal output.
On The Memory HiLogger will sound an audible
beep during alarm signal output.
12
: On
: Off
34
Configuring Alarm Output
Select the [CH] Screen.
To stop the alarm beeper
while measuring, set the
beeper to [Off] in the
ALM channel settings on
the Waveform screen.