3.4 Setting Configuration 2 – Input Channel Settings
53
3
3
Chapter 3 Setting before Measurement
Configure these settings for each pulse channel receiving input from a pulse
output device such as a watt-hour or flow meter.
Use the scaling function to display integrat
ed pulse counts as the correspond-
ing physical units (such as Wh or VA) of the measurement parameter (p. 58).
Integration (Count) Measurement Settings
When using the scaling function, the integrated pulse count can
be displayed after conversion to the measurement target’s prop
-
erty (Wh, VA, etc.). (p. 58)
The count will be saturated when the pulse count exceeds
1,073,741
,823. If there is a possibility that this number may be
exceeded, it is recommended to set the integration mode to
instantaneous and tabulate the results later with Excel or a simi-
lar application.
1 Select a channel (P 1 to P 4) for the [Chan-
nel]
setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Count] for the [Input] (input type)
setting.
4 Select the measurement method from the
[Count Mode] settings.
Setting options:( default setting)
5 Select the basis of measurement from the
[Slope] settings.
Setting options:( default setting)
Add* Counts the number of pulses acquired
from the start of measurement.
Inst Counts the number of pulses acquired dur-
ing each recording interval. The pulse
coun
t is reset at the beginning of each re-
cording interval.
Counts the number of L - H transitions.
Counts the number of H - L transitions.
Totalized
[Add]
Instantaneous
[Inst]
0 V
L
H
12
: On
: Off
345