3.4 Setting Configuration 2 – Input Channel Settings
53
3
3
Chapter 3 Setting before Measurement
Configure these settings for each pulse channel receiving input from a pulse
output device such as a watt-hour or flow meter.
Use the scaling function to display integrat
ed pulse counts as the correspond-
ing physical units (such as Wh or VA) of the measurement parameter (p. 58).
Integration (Count) Measurement Settings
• When using the scaling function, the integrated pulse count can
be displayed after conversion to the measurement target’s prop
-
erty (Wh, VA, etc.). (p. 58)
• The count will be saturated when the pulse count exceeds
1,073,741
,823. If there is a possibility that this number may be
exceeded, it is recommended to set the integration mode to
instantaneous and tabulate the results later with Excel or a simi-
lar application.
1 Select a channel (P 1 to P 4) for the [Chan-
nel]
setting contents.
2 To enable measurement on the channel,
select On to display the check mark ( ).
3 Select [Count] for the [Input] (input type)
setting.
4 Select the measurement method from the
[Count Mode] settings.
Setting options:( default setting)
5 Select the basis of measurement from the
[Slope] settings.
Setting options:( default setting)
Add* Counts the number of pulses acquired
from the start of measurement.
Inst Counts the number of pulses acquired dur-
ing each recording interval. The pulse
coun
t is reset at the beginning of each re-
cording interval.
Counts the number of L - H transitions.
Counts the number of H - L transitions.
Totalized
[Add]
Instantaneous
[Inst]
0 V
L
H
12
: On
: Off
345