9.1 Connecting to the External Control Terminals
130
The external control ground terminal is not isolated from the
Memory HiLogger’s chassis ground. Make certain that there will
be no potential difference between the external control ground
terminal and the ground of any connected device. Otherwise, the
Memory HiLogger or device could be damaged.
To avoid electric shock, use the recommen
ded
wire type to con-
nect to the external control terminals, or otherwise ensure that
the w
ire used has adequate current-handling capacity and insu-
lation.
Terminal Connections
Required items:
Recommended cables:
Single strand diameter
0.65 mm (AWG22),
Multi-strand
0.32 mm
2
(AWG22)
Usable cables:
Single strand diameter
0.32 to 0.65 mm (AWG28 to 22),
Multi-strand 0.08 to 0.32 mm
2
(AWG28 to 22)
Strand diameter
0.12 mm or greater (per wire)
Standard insulation stripping length: 9 to 10 mm
Button operation specified tool: Flat-blade screwdriver
(shaft diameter
3 mm, tip width 2.6 mm)
Single strand
Multi-strand
10 mm
1 Push in the button on the connector with a flat-
blade screwdriver or other tool.
2 With the button held in, insert the cable into the
cable connection hole.
3 Release the button.
The cable is locked.
1
2
3
Top View