Chapter 10 Specifications
143
3
10
Chapter 10 Specifications
Hybrid display
configurations
Simultaneous waveform and gauge display
Waveform display
Simultaneous waveform and numerical value display
Simultaneous numerical value and comment display
Numerical value display
Numerical calculation display (simultaneous calculation results
and wave
form
s)
Cursor value display (simultaneous A/B cursor values and wave-
forms)
Waveform
compression and
magnification
Time axis (horizontal axis)
100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s, 1min, 2min,
5min, 10min
, 20min, 30min, 1h, 2h, 5h, 10h, 12h, 1day/division
Voltage axis (vertical axis)
×100, ×50, ×20, ×10, ×5, ×2, ×1, ×1/2
(switchable to upper and lower limit settings)
Waveform scrolling Horizontal scrolling
Waveform storage The last 3.5 M data points (recording time for one analog channel)
are stored in internal memory.
Data retained in memory can be observed by backward scrolling.
Basic Specifications
Trigger Section
Trigger method Digital comparison
Trigger timing Start, stop, start&stop
Trigger source Analog input (CH1 to 10), Integrated pulse input (P1 to 4)
External trigger Triggering occurs when the external trigger and ground terminals
are shorted, or on falling edges from High (3 to 5 V) to Low (0 to 0.8
V) levels.
Maximum input voltage: 0 to 7 V DC
Minimum response pulse width: 1 ms H, 2
s L
Trigger criteria AND or OR of each trigger source, Trigger criteria can be set for
each channel.
Trigger types
(analog and pulse)
Level Trigger
Triggering occurs when the signal rises (or falls) through a
specified value.
Window Trigger
Triggering occurs when the input signal level enters or exits the
amplitude window defined by upper/lower thresholds.
Trigger level
resolution
Analog channel: 0.025%f.s. (f.s. = 10 divisions)
Pulse channel: 1c (Pulse count integration), 1/n r/s (revolution)
n = pulses/revolution
Pre-trigger Time span setting (any span). Settable with real-time saving.