Loading...
Operating Precautions
11
Using a CF Card/USB flash drive
Inserting a CF card/USB flash drive upside down, backwards or in the
wrong direction may damage the CF card, USB flash drive, or HiLogger.
Never eject a CF card /USB flash drive while measuring or when the
HiLogger
is or accessing the card. Data on the CF card/USB flash
drive may be destroyed. (The CF icon/USB flash drive icon at the
lower right is red while the card is being accessed.)
Do not transport the HiLogger while a USB flash drive is connected.
Dam
age could result.
As the CF card/USB flash drive is sensitive to static electricity, damage
to the CF card/USB flash drive or wrong operations by the HiLogger
may occur due to static electricity. Please be careful when handling it.
With some USB flash drives, the HiLogger may not start up if power is
turned
on while the USB flash drive is inserted. In such a case, turn
power on first, and then insert the USB flash drive. It is recommended
to try out operation with a USB flash drive before starting to use it for
actual measurements.
The Flash memory in a CF card/USB flash drive has a limited operat-
ing life. After long-term usage, data storage and retrieval become diffi-
cult. In this case, replace the CF card/USB flash drive with a new one.
We cannot provide compensation for data loss in a CF card/USB flash
dr
ive, regardless of content or cause of the damage. Data is also cleared
from memory if a long time passes after measuring. Always maintain a
backup of important data stored on a CF card/USB flash drive.
Although real-time saving to USB flash drive is supported, a CF card is
r
ec
ommended for data preservation. Performance cannot be guaranteed
when using storage media other than a Hioki-specified CF card option.
Use a USB flash drive whose continuous current consumption does not
exce
ed 300 mA (peak 500 mA). (The peak value is displayed as "Max
Power" under the USB flash drive self-test on the [System] screen.)
Depending on how USB is used, the USB connector and instrument set-
tings may vary as shown in the chart below.
The three USB methods of use described in the chart below involve
exclu
sive settings and cannot be used simultaneously.
USB method of use
Connector
used
[System] screen USB
mode setting
Use a USB flash drive. Type A USB Memory (Default)
Communicate with the LR8431-20
and initiate measurement using the
Logger Utility software from a com-
puter (using a USB cable).
Type B USB Communication
Read files on a CF card that is con-
nected to the LR8431-20 from a
computer (using a USB cable).
Type B USB Drive