3.3 Setting Configuration 1 – Recording Settings
47
3
3
Chapter 3 Setting before Measurement
1 Select which content to save from the [Auto
Save] settings.
Setting options: Off*, Waveform(realtime), CSV(re-
altime), Calc(post
meas.), Waveform + Calc, CSV
+ Calc ( default setting)
When [Waveform(realtime)], [CSV(realtime)],
[Waveform + Calc], [CSV + Calc] is selected
2 (As needed) You can enter a file name for the
waveform files.
See: 3.6, "Entering Comments" (p. 60)
If left blank, file names are created automatically.
See: Appendix 2, "File Naming" (p. A7)
3 Select the removable storage priority (CF card
or USB flash drive)
4 Select the [Deleting] method for saving data when
the CF card or USB flash drive becomes full.
Setting options:( default setting)
5 Select whether to create a new file for each
recording length, or to create new files at
specified intervals (using
[Split Save]).
Setting options:( default setting)
Off* Saving stops when the CF card or USB flash
drive beco
mes full.
On When space on the CF card or USB flash drive
de
cli
nes to a certain level when starting or dur-
ing automatic saving, the oldest file is deleted.
If the
file cannot be deleted, saving stops. Only
waveform files can be deleted in this way.
Appendix 2, "File Naming" (p. A7)
Off* One file is created for each recording
length.
On Specify the length (time span) of saved
s
egme
nts. A measurement data file is cre-
ated for each segment.
Ref TimeSpecify a reference time and segment
le
ngth (time span). A file is saved after
each segment time span beginning at the
reference time.
1
2
3
4
[Split Save: On]
[Split Save: Ref Time]
Ref Time (reference time):
0:0 to 23:59
Length (Segment length):
1, 2, 5, 10, 15, 20, 30 minutes,
1, 2, 3, 4, 6, 8, 12 hours, 1day
Segment length: Days (0 to
30), Hours (0 to 23), Minutes
(0 to 59)
When starting measurement,
the message “Sync fixed
time” is displayed until the re-
cording interval synchronizes
with specified reference time.
Horizontal axis settings are in-
hibited during automatic sav-
ing of waveforms (p. 43).
5
Regardless of the split save setting, the file will be automatically
split when the file size nears 2 GB.