9.4 Saving, Display and Deleting Measurement Data
165
89
Chapter 9 Data Saving and File Operations
(SYSTEM-MEMORY screen)
9
When you move the cursor, on the [SYSTEM]-[Memory]-[List] screen, to a data storage folder you wish to
display, the F1 [Load] key will appear.
When you press the F1 [
Load] key, [Analyze] will be activated, displaying the event list, trend data, detailed
trend data in the specified folder.
Event, trend data and detailed trend data can be checked.
Refer to "Chapter 8 Checking Events (EVENT screen)" (p.139) for the Event Confirmation Method.
Return to [
Settings] with the DATARESET key.
[TIME PLOT]
screen maximum display times
Display Measurement Data (Load)
• The maximum displayed times of the trend data, detailed trend data, and har-
monic trend data in the
[TIME PLOT] screen of the Hioki PQ3198 is subject to
certain constraints. To confirm all measured trend data, use the application
software PQ ONE, which is supplied with the instrument.
• Data measured in different versions will not load ev
en if the instrument is the
same.
• The F1 [
Load] key will appear when the cursor is in the stored data folder.
(********).
TIME PLOT
Interval
Recording Items setting
All data
(Saves all data)
Power and Harmonic
(Saves RMS values and harmonics)
Power
(Saves RMS values only)
1 second 7 min. 52 sec. 15 min. 44 sec. 2 hours 37 min. 20 sec.
3 seconds 23 min. 36 sec. 47 min. 12 sec.
7 hours 52 min.
15 seconds 1 hour 58 min. 3 hours 56 min. 1 day 15 hours 20 min.
30 seconds 3 hours 56 min. 7 hours 52 min.
3 days 6 hours 40 min.
1 minute 7 hours 58 min. 15 hours 44 min. 6 days 13 hours 20 min.
5 minutes 1 day 15 hours 20 min. 3 days 6 hours 40 min. 32 days 18 hours 40 min.
10 minutes 3 days 6 hours 40 min. 6 days 13 hours 20 min. 35 days
15 minutes 4 days 22 hours 9 days 20 hours
35 days
30 minutes 9 days 20 hours 19 days 16 hours 35 days
1 hour 19 days 16 hours 35 days
35 days
2 hours 35 days 35 days 35 days
150/180 wave
(Approx. 3 sec)
23 min. 36 sec. 47 min. 12 sec. 7 hours 52 min.