3.8 Turning the Power On and Off (Setting the Default Language)
48
Be sure to read the "Usage Notes" (p.7) before turning the instrument on.
Turn on the instrument after connecting the AC adapter, voltage cords, and current sensors.
For best precision, allow at least 30 minutes warm-up be
fore executing zero adjustment and measuring.
3.8 Turning the Power On and Off (Setting the
Default Language)
Turning the power on
Turn the POWER switch on ( | ).
The instrument performs a 10-second power-on self test.
See: 3.3 (p.42)
After the self-test is complete, the [SYSTEM]-[Wiring] screen will be displayed.
If the self-test fails, operation stops at the self-test screen. If the fault recurs after turn-
ing the power off and on, the instrument may be
damaged. Perform the following proce-
dure:
1. Cancel measurement and disconnect the voltag
e cords and current sensors from
the measurement line before turning off the instrument's POWER switch.
2. Disconnect the power cord, voltage cords, an
d current sensors from the instrument.
3. Contact your authorized Hioki distributor or reseller.
Turning the power off
Turn the POWER switch off ( ).
After use, always turn OFF the power.
Do not turn the instrument off with the voltage cords and current sensors connected to
the measurement line. Doing so may damage the instrument.