13.1 General Specifications
191
10
13
Chapter 13 Specifications
Specifications Chapter 13
13.1 General Specifications
Operating environment Indoor use, pollution degree 2, altitude up to 3000 m (9843 ft.)
At an altitude of above 2000 m (6562 ft.), the measurement categories are lowered to 600 V CAT III.
Operating temperature and humidity 0°C to 30°C (32°F to 86°F) 95% RH or less (no condensation)
When charging battery: 10°C to 30°C (50°F to 86°F)
30°C to 50°C (86°F to 122°F) 80% RH or less (no condensation)
When charging battery: 30°C to 35°C (86°F to 95°F)
Storage temperature and humidity -20°C to 30°C (-4°F to 86°F) 95% RH or less (no condensation)
30°C to 50°C (86°F to 122°F) 80% RH or less (no condensation)
(If the instrument will not be used for an extended period of time, remove the battery pack and store
in a cool location [from -20°C to 30°C (-4°F to 86°F)].)
Dust and water resistance IP30 (EN60529)
Standards Safety EN61010
EMC EN61326 Class A
Power supply quality
measurement method
IEC 61000-4-30 Ed2:2008 Class A, IEEE 1159
Power supply Z1002 AC Adapter
Rated supply voltage : 100 V AC to 240 V AC (Voltage fluctuations of
±10% from the rat-
ed supply voltage are taken into account.)
Rated power supply frequency : 50 Hz/60 Hz
Anticipated transient overvoltage: 2500 V
Maximum rated power : 80 VA (When charging, including the AC adapter)
35 VA (When charging, main unit only)
Z1003 Battery Pack
Rated supply voltage : 7.2 V DC
Maximum rated power : 8 VA
Recharge function Charges the battery regardless of whether the instrument is on or off.
Charging time 5 hours 30 minutes at a maximum (at 23°C, as a referential)
Continuous operating time When Z1003 Battery Pack is used (at 23°C, as a referential)
Approx. 3 hours (fully charged, continuous operation, LCD backlight AUTO OFF)
Backup battery Approx. 10 years (at 23°C, as a referential)
For backup clock and setting conditions (Lithium battery)
Maximum recording interval Repeat recording function, 1 week: 55 weeks
Repeat recording function, 1 day: 366 days
Repeat recording function, off: 35 days
Maximum recording events 9999 events (Switchable between 1000 events and 9999 events)
Clock function Auto-calendar, leap-year correcting 24-hour clock
Real-time clock accuracy ±0.3 s per day (instrument on, 23°C±5°C [73°F±9°F])
±1 s per day (instrument on, within operating temperature and humidity range)
±3 s per day (instrument off) (at 23°C, as a referential)
Display refresh rate Approx. 0.5 s
Display 6.5-inch TFT color LCD (640 × 480 dots)
Interface SD memory card, USB, LAN, RS-232C, External I/O
Dimensions Approx. 300W × 211H × 68D mm (11.81″W × 8.31″H × 2.68″D)
Body Strap can be attached.
Mass Approx. 2.5 kg (88.2 oz.) (with Z1003 Battery Pack installed)
Product warranty period 3 years
Accessories See: "Accessories" (p.4)
Options See: "Options" (p.5)