12.3 Remote Control of the Instrument by Internet Browser
186
Click on the control panel keys to perform the same operations as the instrument keys.
To enable automatic browser screen updating, set th
e update time in the auto update menu.
Auto display
update
The instrument screen emulation updates at the specified interval.
Auto display update
Control panel emulation
Instrument screen emulation
Setting Contents:( ∗ : Default setting)
OFF, 0.5∗/ 1/ 2/ 5/ 10 sec
The displayed information may vary with the browser being used.
If the instrument does not accept key input
Is the browser's security level set to “High”, or has JavaScript been disabled?
Change the browser's security setting to Medium or Medium-high.