13.3 Screen Specifications
218
(5) Event Settings
Event hysteresis 0% to 10% (Applies to all parameters except frequency.)
Fixed to 0.1 Hz for frequency; percentage of threshold value for other parameters.
Maximum recordable events 1000/9999
Sets the maximum number of recordable events per measurement
when repeat recording is off.
When the repeat recording function is on, the number of events is obtained by multiplying this set-
ting by the repeat count. A setting of 9999 disables voltage waveform comparison events.
Slide reference voltage OFF/ON (Applies to swells and dips.)
When turned on, the slide reference voltage is used instead of the nominal voltage.
Timer event count OFF/1 minute/5 minutes/10 minutes/30 minutes/1 hour/2 hours
Events are generated at the chosen interval.
Continuous event count OFF/1/2/3/4/5 times
Applies to all events being recorded. When time targe
t events occur, they are automatically treated
as sequential events if the event in question occurs the set number of times. However, events oc-
curring during sequential events cannot trigger sequential even
ts. In addition, generation of se-
quential events stops when recording stops.
External event OFF,ON
Event setting details See: "5.6 Changing Event Settings" (p.85)
(6) [MEMORY] Screen
Target interface SD memory card
Function Mass storage, saving (of settings data), loading (of settings data, measurement data, event data,
screen data, and version upgrade files), deletion of folders and files, and formatting