13.8 Calculation Formula
237
10
13
Chapter 13 Specifications
-7. Harmonic Voltage (Uharm), Harmonic Current (Iharm), Inter-harmonic voltage
(Uiharm), Inter-harmonic current (Iiharm)
Single Phase 2-wire
1P2W
Single
Phase
3-wire
1P3W
3-Phase,
3-Wire,
2-Measure-
ment
3P3W2M
3-Phase,
3-Wire,
3-Measure-
ment
3P3W3M
3-Phase,
4-Wire
3P4W
Uharm[Vrms]=Uck
(including adjacent inter-harmonic com-
ponents)
U
1k
U
4k
U’ck=
Uck=
U
1k
U
2k
U
4k
U
12k
U
32k
U
4k
U
12k
U
23k
U
31k
U
4k
U
1k
U
2k
U
3k
U
4k
For 3-phase 3-wire connections, indicates the result of harmonic calculations using the
line voltage. For 3-phase 4-wire connections, indicates the result of harmonic calculations
using the phase voltage.
The harmonic voltage content percentage is calcula
ted by dividing the harmonic voltage
component for the specified order by the fundamental voltage component and multiplying
by 100.
When k = 0, the Uc0 component is treated as DC for order 0.
For 60 Hz measurement, the value 10 in the formula is
replaced with 12. For 400 Hz mea-
surement, the value 10 in the formula is replaced with 80.
Iharm[Arms]=Ick
(including adjacent inter-harmonic com-
ponents)
I
1k
I
4k
I’ck=
Ick=
I
1k
I
2k
I
4k
I
1k
I
2k
I
4k
I
1k
I
2k
I
3k
I
4k
I
1k
I
2k
I
3k
I
4k
The harmonic current content percentage is calculated by dividing the harmonic current
component for the specified order by the fundamental current component and multiplying
by 100.
When k = 0, the Ic0 component is treated as DC for order 0.
When using 60 Hz, the number "10" in the expression
above is "12." When using 400 Hz,
the number "10" in the expression above is "80."
Uiharm[Vrms]=Uck U
1k
U
4k
U’ck=
Uck=
U
1k
U
2k
U
4k
U
12k
U
32k
U
4k
U
12k
U
23k
U
31k
U
4k
U
1k
U
2k
U
3k
U
4k
The values 3 and -3 in the formula apply to 50 Hz measurement and are replaced with 4
and -4 for 60 Hz measurement. In the formula, k = 0.5, 1.5, 2.5, 3.5,...
For 3-phase 3-wire connections, indica
tes the result of harmonic calculations using the
line voltage. For 3-phase 4-wire connections, indicates the result of harmonic calculations
using the phase voltage.
The inter-harmonic voltage content percentage is calcu
lated by dividing the inter-harmonic
voltage component for the specified order by the fundamental voltage component and
multiplying by 100.
For 60 Hz measurement, the value 10 in the formula is replaced with 12.
Phase System
Items
Uckr()
2
Ucki()
2
+{}
U
2
c10k n+()10()
n
1=
1
Ickr()
2
Icki()
2
+{}
I
2
c10k n+()10()
n
1=
1
Uckr()
2
Ucki()
2
+{}
U
2
c10k n+()10()
n
3=
3
HIOKI PQ3198A961-02