5.6 Changing Event Settings
91
Chapter 5 Changing Settings (as necessary)
5
Events can be configured by pressing the DF3 key to display the [Power/etc] screen. Events are gener-
ated at the set interval and recorded as external events.
Once recording is started, timer events
will be recorded at a fixed interval (the set time) from the start
time.
A function to continuously generate the number of set e
vents (1 time to 5 times) automatically each time
an event is generated.
Event apart from the first event will be recorded
as "continuous event".
Due to this, instantaneous waveforms of up to one second can be recorded after an event has been gen-
erated.
However, in an event generated during a c
ontinuous event generation, continuous event will not be gen-
erated.
Continuous event generation
will stop as soon as the measurement is finished.
Used to observe the instance when the event is g
enerated and the changes in the instantaneous wave-
forms after that. In the case of this inst
rument, waveforms of up to one second are recorded.
Generating events periodically (timer event settings)
[Timer Event]
Accept setting
Cancel
Set the interval at
which to generate
the timer event
Select the setting
to change it
Event recorded
Ext
Event recorded
Ext
Event recorded
Ext
Event recorded
Ext
Recording
started
Interval Interval Interval Interval
Generating Events Continuously (Continuous Event Function)