Command Reference
:MOHM:RANGe
Sets the resistance range.
Syntax
:MOHM:RANGe <data>
<data> 2 M / 20 M / 200 M / 2000 M / 4000 M / AUTO (text data)
Example To set to 200 M
Ω
range
PC>
:MOHM:RANGe 200M
Error An execution error occurs if the 2000 M
Ω
range is set when the test
voltage is 25 V to 99 V.
:MOHM:RANGe?
Queries the resistance range. The resistance range is returned in text data (<data>).
Syntax
:MOHM:RANGe?
Response
:MOHM:RANGE <data>
or simply
<data>
<data> 2 M / 20 M / 200 M / 2000 M / 4000 M / AUTO (text data)
Example To query the resistance range (when the resistance range is 200 M
Ω
)
PC>
:MOHM:RANGe?
ST5520>
:MOHM:RANGE 200M
(when the
:HEADer
command is ON)
ST5520>
200M
(when the
:HEADer
command is OFF)
:SPEed
Sets the measurement speed to FAST/SLOW.
Syntax
:SPEed <data>
<data> FAST/SLOW (text data)
Example To set the measurement speed to FAST
PC>
:SPEed FAST
:SPEed?
Queries the measurement speed. The measurement speed setting FAST/SLOW is returned in
text data (<data>).
Syntax
:SPEed?
Response
SPEED <data>
or simply
<data>
<data> FAST/SLOW (text data)
Example To query the measurement speed (when the speed is set to FAST)
PC>
:SPEed?
ST5520>
:SPEED FAST
(when the
:HEADer
command is ON)
ST5520>
FAST
(when the
:HEADer
command is OFF)