9 - 23
Cygnus 2 Operating Manual
9.4.10 Query Source Parameter
Command = QC <Command ID> <Source number>
Command ID = <Byte> See Table 9-8.
Source Number = <Byte> Source number 1 - 6
Response = <Integer> | <Float>
Type of response depends on command ID, see Data Type column in Table 9-8.
9.4.11 Update Source Parameter
Command = UC <Command ID> <Source number> <Parameter Value>
Command ID = <Byte> See Table 9-8.
Source Number = <Byte> Source number 1 - 6
Parameter Value = <Integer> | <Float>
Type of value depends on command ID, see Data Type column in Table 9-8.
Response = None (Just header and trailer)
5
(0x05)
# of Positions 1/ 12/ 1 Integer
Visible only when Sensor
Type = 4
Update not allowed if
the Sensor is running.
6
(0x06)
# of Pulses 1/ 10/ 1 Integer
7
(0x07)
Pulse On (Time)
(Seconds)
0.1/ 9.9 /1 REAL_3X1
8
(0x08)
Pulse Off (Time)
(Seconds)
0.1/ 9.9 /1 REAL_3X1
9
(0x09)
Recorder Output 0/ 12/ 0 Integer 0 = not defined Not allowed to be the
same as a Source DAC
10
(0x0a)
Recorder Function 0/ 2/ 0 Encode 0 = Unfiltered Rate
1 = Thick 2 = RateDev
Note: Rate is not
affected by Rate Filter
Time
11
(0x0b)
Recorder Range 0/
99999/
100
Integer Gray font when
Recorder Function =
Rate Deviation
Table 9-7 Sensor parameters (continued)
QS
&
US
Cmnd
ID
Name
(Units)
Range:
Low Limit/
High Limit/
Default
Data Type
Notes
Update Restrictions
and Display Notes