6 - 8
Cygnus 2 Operating Manual
6.8 LOCK Page Code Set-up
Figure 6-7 General lock page
PROGRAM LOCK CODE . . . . . . . . 1 to 9999
The lock code is set by moving to Program Lock Code and entering the desired
code of up to 4 digits. If no lock code is desired, no entry needs to be made.
The lock code will be in effect after turning the unit off and an again.The text
"L LOCK" will be displayed when Program Lock is set. Default value is 0.
NOTE: If a lock code was previously programmed, that code must
subsequently be entered on the Main Menu screen, with the cursor
positioned on XXXX at the Lock Code XXXX at the lower right corner
of the Main Menu options area. After entering the lock code on that
screen, the lock code can be cleared if desired by entering 0 for the
Program Lock Code on the General / Lock page.The lock code can
also be cleared by pressing CLEAR on power-up, however, if no lock
code has been programmed, this will reset all parameters to their
default values.
NOTE: Program lock codes are not stored on and retrieved from the USB port.