13 - 25
Cygnus 2 Operating Manual
13.7.6 Crystal Snatcher
Use the crystal snatcher, supplied with the sensor, as follows:
1 Insert crystal snatcher into ceramic retainer (1) and apply a small amount of
pressure. This locks the retainer to the snatcher and allows the retainer to be
pulled straight out (2). See Figure 13-6.
2 Reinsert the retainer into the holder after the crystal has been replaced.
3 Release the crystal snatcher with a slight side-to-side motion.
Figure 13-6 Use of the crystal snatcher
13.7.7 CrystalSix (XtalSix)
Refer to the Crystal Six Operating Manual (IPN 074-155) for specific instructions
for this device.