9 - 27
Cygnus 2 Operating Manual
9.4.19 Update Output Name
Command = UO <Output Number> <Output Name>
Output Number = <Byte> 1 - 38
Output Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
Output Names may only be updated if Cygnus 2 is in Ready or Stop.
NOTE: UO changes are not allowed while a Material is running. The UO command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name
9.4.20 Query Output Type
Command = QT <Output Number>
Output Number = <Byte> 1 - 38
Response = <Output Type>
Output Type = <Byte> 0 or 1 (0 = NO, 1 = NC)
9.4.21 Update Output Type
Command = UT <Output Number> <Output Type>
Output Number = <Byte> 1 - 38
Output Type = <Byte> 0 or 1 (0 = NO, 1 = NC)
Response: None (Just header and trailer)
NOTE: UT changes are not allowed while a Material is running.
9.4.22 Query User Message
Command = QV <User Message Number>
User Message Number = <Byte> 1 - 10
Response = < User Message>
User Message = <String> Up to 19 characters, null terminated