EasyManuals Logo

Inficon Cygnus 2 User Manual

Inficon Cygnus 2
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
9 - 26
Cygnus 2 Operating Manual
9.4.14 Query Material Name
Command = QN <Material Number>
Material Number = <Byte> 1-6
Response = < Material Name>
Material Name = <String> Up to 15 characters, null terminated. The string returned
will be a blank if the default Material # name has not been changed.
9.4.15 Update Material Name
Command = UN < Material Number> < Material Name>
Material Number = <Byte> 1 - 6
Material Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
9.4.16 Query Input Name
Command = QI <Input Number>
Input Number = <Byte> 1 - 28
Response = <Input Name>
Input Name = <String> Up to 15 characters, null terminated
9.4.17 Update Input Name
Command = UI <Input Number> <Input Name>
Input Number = <Byte> 1 - 28
Input Name = <String> Up to 15 characters, null terminated
Response = None (Just header and trailer)
Input Names may only be updated if Cygnus 2 is in Ready or Stop.
NOTE: UI changes are not allowed while a Material is running. The UI command
will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name.
9.4.18 Query Output Name
Command = QO <Output Number>
Output Number = <Byte> 1 - 38
Response = <Output Name>
Output Name = <String> Up to 15 characters, null terminated

Table of Contents