13 - 7
Cygnus 2 Operating Manual
Datalog Failure Cygnus 2 is sending the Datalog
information faster then the receiving device
can accept the data. The data is lost.
The message is
removed from
the screen on
any key press or
if another
transient
message
appears
All 11
Ethernet In Use Ethernet port is active and the Ethernet
parameters can’t be changed. The
command sent to the Ethernet card did not
go through or the Ethernet card did not
send a valid message indicating the update
of values was completed.
General 12
No Rotate: CrystalTwo Cannot rotate a CrystalTwo. Sensor Info 13
No Rotate: Layer
Running
Cannot rotate a Sensor while a Material is
running.
Sensor Info 14
Invalid Equation An attempt was made to save an invalid
logic statement. Logic statements must
have connectors and any required numeric
before exiting Logic editing.
Logic edit 15
Power Not Zero A Source maintenance crucible switch was
requested when the Source power is not
zero.
Maintenance 16
RS232 Test Failed RS-232 loop back test failed. General/Comm 17
RS232 Test Passed RS-232 loop back test passed. General/Comm 18
XIU Test Failed XIU test failed Sensor Info/Type 19
XIU Test Passed XIU test passed. Sensor Info/Type 20
No XIU Test: Switching Cannot test an XIU while a crystal switch is
in process.
Maintenance 21
Edit In Process When an edit is process and the menu key
is pressed.
All 22
USB Disk Error Fault Conditions: No Disk, Folder/File
Read Only, Firmware Invalid, Firmware No
Upgrade, Exceeded max. file count
Storage 23
USB error: Disk Full Disk is Full Storage 24
Table 13-4 Transient messages (continued)
Transient Message Set Clear Screen Remote
Comm.
Encodes