ULTRAFLOW®54(H)/(J)
70
KamstrupA/S∙TechnicalDescription∙5512‐1554_J1_GB_04.2018
10.4 Application
Flow sensor adjustment.
Before adjusting a sensor you must make sure that the sensor operates satisfactorily in the flow rig in question.
See paragraph 9 Calibrating ULTRAFLOW
®
.
If it is necessary to adjust the sensor more than a few per cent, the sensor is probably defective, or has a different
error, and should not be adjusted.
10.4.1 Choice of COM-port
Open ”Setup”
Select a COM-port for ULTRAFLOW
®
X4.
The USB driver must be installed before connecting
the interface.
The related COM-port will not appear from the list until
the USB Interface has been connected.
Select COM-Port for Pulse Divider.
Activate ”Save” in order to save the selected ports.