ULTRAFLOW®54(H)/(J)
48
KamstrupA/S∙TechnicalDescription∙5512‐1554_J1_GB_04.2018
-6
-4
-2
0
2
4
6
0,0 1 0,1 1 10
Tolerances [%]
Flow [m
3
/h]
Flow sensor tolerances q
i
:q
p
1:100 (q
p
1.5 m
3
/h)
EN1434 kl.3
EN1434 kl.2
½ EN1434 kl.2
q
s
q
p
0.1x q
p
q
i
Figure 43. Flow sensor tolerances q
i
:q
p
1:100 for q
p
1.5 m³/h.
8.9 Power consumption
The current consumption of ULTRAFLOW
®
is as follows:
Max. average 50 μA
Max. current 7 mA (max. 40 ms)
8.10 Interface connector/serial data
ULTRAFLOW
®
54 is fitted with a four-pole connector under the cover. Thus, it is not possible to access the connector
without breaking the seal. The cover is supplied with a factory seal, in connection with verified sensors it will be a
laboratory seal (legal seal).
The connector is used for:
Meter programming, including adjustment of flow charts by means of METERTOOL
Bringing the sensor into test mode
Reading accumulated water quantity in connection with calibration
External control of start/stop in connection with calibration
Due to different physical print layouts the interface connectors of the flow sensors q
p
0.6…2.5 m
3
/h (Type (H)) are
slightly different from the ones of the flow sensors q
p
≥ 3.5 m
3
/h (Type (J)). The interface connectors are constructed
as shown in Figure 44. Note, that the general function of the connectors is identical, but the positioning of single
functions is different.