2600AS-901-01 Rev. B / September 2008 Return to Section Topics 19-81
Series 2600A System SourceMeter® Instruments Reference Manual Section 19: Remote Commands
lan.trigger[N].connected
Replace N with values 1-8.
Attribute The LAN event connection state.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage connected = lan.trigger[lanevent].connected
lanevent The LAN event number.
connected The LAN event connection state.
Remarks •Set to true when the LAN trigger is connected and ready to send trigger events following a
successful lan.trigger[n].connect command. If the LAN trigger is not ready to send
trigger events, this value is set to false.
•Set to false when either lan.trigger[n].protocol or lan.trigger[n].ipaddress
attributes are changed or the remote connection closes the connection.
lan.trigger[N].disconnect
Replace N with values 1-8.
Function Disconnects the LAN trigger.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage lan.trigger[lanevent].disconnect()
lanevent The LAN event number.
Remarks • For TCP connections, this closes the TCP connection.
• The LAN trigger will automatically disconnect when either the lan.trigger[n].protocol or
lan.trigger[n].ipaddress attributes for this event are changed.
lan.trigger[N].EVENT_ID
Replace N with values 1-8.
Attribute Indicates the trigger event detector LAN event number.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage event_id = lan.trigger[lanevent].EVENT_ID
event_id The trigger event number.
lanevent The LAN event number.
Remarks • Set the stimulus of any trigger event detector to the value of this constant to have it respond to
incoming LAN trigger packets.