2600AS-901-01 Rev. B / September 2008 Return to Section Topics 19-141
Series 2600A System SourceMeter® Instruments Reference Manual Section 19: Remote Commands
smuX.trigger.arm.set
X = SMU channel (a or b)
Function Sets the arm event detector to the detected state.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage smuX.trigger.arm.set()
Remarks The SMU will automatically clear all the event detectors when the smuX.trigger.initiate
function is executed. This function should be called after the sweep is initiated.
Also see smuX.trigger.initiate
smuX.trigger.arm.stimulus
X = SMU channel (a or b)
Attribute Selects which event will cause the arm event detector to enter the detected state.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage eventid = smuX.trigger.arm.stimulus
smuX.trigger.arm.stimulus = eventid
eventid Event that triggers the arm detector.
Remarks Set this attribute to zero to bypass waiting for an event.
Set this attribute to the event ID of any trigger event generator to wait for that event.
smuX.trigger.ARMED_EVENT_ID
X = SMU channel (a or b)
Attribute The armed event number.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage event_id = smuX.trigger.ARMED_EVENT_ID
event_id Armed event number.
Remarks Set the stimulus of any trigger event detector to the value of this constant to have it respond to
armed events from this SMU.