19-218 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
tsplink.trigger[N].stimulus
Replace N with the number of the synchronization line: 1-3.
Attribute Selects the event that causes the synchronization line to assert a trigger.
Default 0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage stimulus = tsplink.trigger[n].stimulus
tsplink.trigger[n].stimulus = stimulus
n The trigger line number.
stimulus The event identifier for the triggering event.
Remarks • This attribute selects which event will cause the synchronization line to assert a trigger.
• Set this attribute to zero to disable automatic trigger assertion on the synchronization line.
• Do not use this attribute when triggering under script control. Use
tsplink.trigger[n].assert instead.
tsplink.trigger[N].wait
Replace N with the number of the synchronization line: 1-3.
Function Waits for a trigger.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage triggered = tsplink.trigger[n].wait(timeout)
timeout Specifies the time-out value in seconds.
triggered A customized variable that stores the value true if a trigger
trigger is detected, or false if a trigger is not detected during
the time-out period.
n Trigger number on which to wait.
Remarks This function waits up to timeout value in seconds for an input trigger. If one or more trigger
events were detected since the last time tsplink.trigger[n].wait or
tsplink.trigger[n].clear was called, this function will return immediately. After waiting for a
trigger with this function, the event detector is automatically rearmed and reset. This functionality is
true regardless of the number of events detected.
Details See Interactive triggering in Section 10.
Also see tsplink.trigger[N].clear

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals