Source-Measure Concepts 5-17
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
V-Source (V
S
)
I-Meter
(I
M
)
5mA
10mA
10V
2k
DUT Load Line (R)
I
M
= V
S
/ R
V
S
= I
M
R
= 10V/2k
= 5mA
A) Normal V-source operation
Current Limit
Load Line
Operating
Point
Voltage Source
Load Line
I-Meter
(I
M
)
10mA
100V
= (10mA) (800)
= 8V
B) V-Source in compliance
80V
800
DUT Load Line (R)
V-Source (V
S
)
Figure 5-8
V-Source operating
examples

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω