SCPI Command Reference 17-5
Command Description
Default
parameter
SCPI
Source
memory
:CALCulate2
:LIMit3
:LOWer
[:DATA] <n>
[:DATA]?
:SOURce2 <NRf>
| <NDN>
:SOURce2?
:PASS
:SOURce2 <NRf>
| <NDN>
:SOURce2?
:STATe <b>
:STATe?
:FAIL?
:LIMit512
:UPPer
[:DATA] <n>
[:DATA]?
:SOURce2 <NRf>
| <NDN>
:SOURce2?
Congure lower limit:
Specify lower limit (-9.999999e20 to
9.999999e20).
Query lower limit.
Specify output fail pattern for grading mode
(0 to 7, 3-bit; 0 to 15, 4-bit).
2
Query fail bit pattern.
1
Path to specify pass pattern for sorting mode:
Specify output pass pattern (0 to 7, 3-bit;
0 to 15, 4-bit).
2
Query pass bit pattern.
1
Enable or disable Limit 3 test.
Query state of Limit 3 test.
Return result of Limit 3 test: 0 (pass) or 1 (fail).
Path to control LIMIT 5 to LIMIT 12 tests (see
Note):
Congure upper limit:
Specify upper limit (-9.999999e20 to
9.999999e20).
Query upper limit.
Specify output fail pattern for grading mode
(0 to 7, 3-bit; 0 to 15, 4-bit).
2
Query fail bit pattern.
1
-1
15 or 7
15 or 7
OFF
1
15 or 7
1
The format (ASCII, hexadecimal, octal, or binary) for the returned value is set by FORMat:SOURce2 <name>.
2
Default based on present digital output size (:SOURce2:BSIZe <n>). 3-bit default is 7. 4-bit default is 15. 16-bit default is 65535.
Note: Use LIMit5 through LIMit12 to control Limit 5 through Limit 12 tests respectively. For example, send :LIM5:STAT ON to
enable Limit 5; send LIM10:FAIL? to return the result of Limit 10.
Table 17-1 (cont.)
CALCulate command summary

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
TypeSub-Femtoamp Remote SourceMeter
Compliance Voltage200 V
Maximum Output Voltage200 V
Maximum Output Current100 mA
Compliance LimitProgrammable
InterfaceGPIB, RS-232
Input Impedance>10^14 Ω