EasyManuals Logo
Home>Keithley>Measuring Instruments>6430

Keithley 6430 User Manual

Keithley 6430
558 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #422 background imageLoading...
Page #422 background image
18-4 Performance Verification
Table 18-1
Recommended verification equipment
Description Manufacturer/Model Specifications
Digital Multimeter
1
Hewlett Packard
HP3458A
DC Voltage
DC Current
1V:
10V:
100V:
1000V:
1mA:
10mA:
100mA:
1mA:
10mA:
100mA:
5.6ppm
4.3ppm
6.3ppm
6.1ppm
55ppm
25ppm
23ppm
20ppm
20ppm
35ppm
Resistance Calibrator
1
Fluke 5450A Resistance 19:
190:
1.9k:
19k:
190k:
1.9M:
19M:
65ppm
10.5ppm
8ppm
7.5ppm
8.5ppm
11.5ppm
30ppm
Electrometer
Calibration Standard
2
Keithley 5156 Resistance 100M
1G
10G
100G
200ppm
300ppm
400ppm
500ppm
Precision Resistors
3
Any suitable Resistance 1T
10T
2,000ppm
5,000ppm
Coax Cables (2)
Shielded Cables (2)
Triax-BNC Adapters
Triax-triax Adapters
Keithley 4801
Keithley CA-18-1
Keithley 7078-TRX-GND
Low-noise
Dual banana
Male/female
Male/male
1
90-day, full-scale DMM accuracy specications of ranges required for various measurement points.
2
Characterization accuracy specications shown for Model 5156.
3
Resistors should be accurately characterized to uncertainty shown and mounted in a shielded enclosure.

Table of Contents

Other manuals for Keithley 6430

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6430 and is the answer not in the manual?

Keithley 6430 Specifications

General IconGeneral
BrandKeithley
Model6430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals