Calibration 19-9
Table 19-3
Mainframe voltage calibration summary
Source range
1
Source voltage Multimeter voltage reading
2
000.2V +200.00mV
+000.00mV
-200.00mV
+000.00mV
___________ mV
___________ mV
___________ mV
___________ mV
002V +2.0000V
+0.0000V
-2.0000V
+0.0000V
___________ V
___________ V
___________ V
___________ V
020V +20.000V
+00.000V
-20.000V
+00.000V
___________ V
___________ V
___________ V
___________ V
200V +200.00V
+000.00V
-200.00V
+000.00V
___________ V
___________ V
___________ V
___________ V
1
Use EDIT and RANGE keys to select source range.
2
Multimeter reading used in corresponding calibration step. See procedure.