Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 101
Preparation for Accurate Measurement (Executing Correction)
Obtaining Correction Data
The table below summarizes how data is set up, depending on the setup at
execution.
Measurement conditions during data measurement
The data for the SHORT correction is measured under the following
measurement conditions:
Measurement range mode: auto range mode
Measurement time mode: 8
Output impedance of measurement signal source: 20
For other settings, such as averaging and trigger delay, the measurement
conditions set up at execution are used for measurement.
Measurement procedure
For more information about the measurement procedure for the data used in
the SHORT correction, refer to “Measuring data for SHORT correction” on
page 47.
Multi
correction
Setup of
measurement
signal
frequency
Data
OFF
120 Hz
120 Hz correction data and 120 Hz cable length
for normal operation
1 kHz
1 kHz correction data and 1 kHz cable length for
normal operation
1 MHz
1 MHz correction data, 1 MHz cable length, and
1 MHz frequency shift for normal operation
ON
120 Hz
120 Hz correction data and 120 Hz cable length
for the channel selected at the time of
multi-correction measurement
1 kHz
1 kHz correction data and 1 kHz cable length for
the channel selected at the time of
multi-correction measurement
1 MHz
1 MHz correction data, 1 MHz cable length and
1 MHz frequency shift for the channel selected at
the time of multi-correction measurement