Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 129
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
Step 8. Enter the limit value using the entry keys. When you enter the value, the
softkey labels change to unit labels (p, n, u, m, x1).
You can clear your selected upper limit value by pressing the CLEAR
softkey.
You can enter the upper limit value using the LOWx(-1) softkey.
You can clear all the limit values and sets the status to OFF for your
selected bin by pressing the CLEAR LINE softkey.
Setting up AUX function
If the secondary parameter limit range is ON, the sorting result varies when it
exceeds the secondary parameter limit range, depending on the ON/OFF state
of the AUX function (Table 7-4).
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Use the cursor keys to select the AUX field.
Step 4. Use the following softkeys:
Table 7-4 Sorting result when measured secondary parameter value exceeds limit
range
Primary parameter sorting
result
AUX BIN function Sorting result
One of BIN1 to BIN9
OFF OUT_OF_BINS
ON AUX_BIN
Not sorted to any BINs No relation OUT_OF_BINS
Softkey Function
ON Turns ON the auxiliary function.
OFF Turns OFF the auxiliary function.