Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 123
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
Setting Up Sorting Judgment Conditions
Clearing (resetting) limit ranges
Clearing the limit ranges returns the settings of the following items to the
factory-default preset values (refer to Table C-1 on page 228).
ON/OFF, lower limit value, and upper limit value of all limit ranges (BIN1 to
BIN9 and secondary parameter limit range)
Limit range designation method
Reference value for the tolerance mode
Execution procedure
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Use the cursor keys to select the BIN field.
Step 4. Press the CLEAR TABLE softkey, to clear the limit ranges.
Selecting a limit range designation method
Two methods can be used to designate the limit ranges for the primary
parameter (BIN1 to BIN9). One is to designate the limit boundary value using
an absolute value (absolute mode) and the other is to designate it using a
relative value (deviation) from the reference value (nominal value) (tolerance
mode). The tolerance mode is further divided into a method to designate the
deviation using an absolute value (absolute tolerance mode) and a method to
designate the deviation using the percentage of the reference value (percent
tolerance mode).
e4981auj0017
Only the absolute mode can be used to designate the secondary parameter
limit range.