228 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Initial Settings
Initial Settings, Settings That Can Be Saved/recalled, Settings That Can Be Backed Up
-
Table C-1 Initial settings, settings that can be saved/recalled, settings that can be backed up
Setting items
Initial settings
(factory settings)
Reset
Save/
Recall
Backup
Front panel key
(:SYST:PRES)
*RST
Measurement parameter
Primary parameter CP ]]^×
Secondary parameter D ]]^×
Measurement signal
Frequency 1 kHz ]]^×
1 MHz frequency shift 0% No effect ] × ^
Level 1 V ]]^×
Signal level compensation
(SLC)
OFF ]]^×
Output mode CONT ]]^×
Measurement range
Auto ON ]]^×
Range setup 100 nF ]]^×
Measurement time mode N 6 ]]^×
Averaging
ON/OFF ON ]]^×
Number of counts 1 ]]^×
Cable length 0 No effect 0 × ^
Analog convergence waiting time setup
1.67 m ]]××
1 m ]]××
270 µ ]]××
Trigger
Triggering No effect × ×
Mode INTernal ]]^×
Trigger delay time 0 ]]^×
Trigger
system
Reset No effect × ×
Initiates No effect × ×
Continuous
activation
ON/OFF
ON ] OFF ^ ×
BNC External trigger slope POSitive No effect ] × ^
OPEN
correction
ON/OFF OFF No effect OFF × ^
Parameter type GB No effect GB × ^
SHORT
correction
ON/OFF OFF No effect OFF × ^
Parameter type RX No effect RX × ^