154 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Using Scanner Interface
Using Multi-correction Function
9-
The procedure to turn ON/OFF the multi-correction function is given below.
Step 1. Press [Meas Setup].
Step 2. Press CORRECTION softkey.
Step 3. Use the cursor keys to select the MULTI field.
Step 4. Use the following softkeys:
Selecting a channel
Making a selection using the front panel
The procedure to select a channel using the front panel keys is given below.
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Use the cursor keys to select the CH field.
Step 4. Use the softkeys or numeric entry keys to enter the channel number. When
data is entered using the numeric keys, the softkeys change to units labels (x1).
Softkey Function
ON Turns ON the multi-correction function.
OFF Turns OFF the multi-correction function.
Softkey Function
INCR++ Increments the channel number in steps of 10.
INCR+ Increments the channel number in steps of 1.
DECR- Decrements the channel number in steps of 1.
DECR-- Decrements the channel number in steps of 10.

Table of Contents

Other manuals for Keysight E4981A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E4981A and is the answer not in the manual?

Keysight E4981A Specifications

General IconGeneral
BrandKeysight
ModelE4981A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals