58 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Selecting Measurement Time
4-
Selecting Measurement Time
For details on the actual measurement time, refer to “Measurement time” on
page 188.
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Use the cursor keys to select the MEAS TIME field.
Step 3. Use the numeric keys or following softkeys:
Selecting Cable Length
Set up the length of the measurement cable to compensate errors due to the
extension of the measurement cable. The E4981A lets you select the length
from 0 m, 1 m, or 2 m, depending on the measurement cable you use.
0 m When you do not use the test lead (in other words, when you
connect the test fixture directly to the UNKNOWN terminal).
1 m When you use the
Keysight 16048A/B Keysight test lead.
2 m When you use the
Keysight 16048D test lead.
Setup procedure
For details on setup procedure, refer to “Setting up cable length” on page 43.
Measurement time of 1, 2, 4, 6 and 8 are available in E4981A.
Softkey Description
INCR+ Increments the measurement time.
DECR- Decrements the measurement time.
Cable length information is saved in internal FLASH memory when power is
OFF with backup function enabled.