Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 63
Setting up Measurement Conditions and Display
Setting Up Display
Setup procedure
Selecting floating/fixed point display
Step 1. Press [Display Format].
Step 2. Use the cursor keys to select the field in which a measurement result of
primary parameter or secondary parameter is displayed.
Step 3. Use the following softkeys:
Displaying measurement result in deviation from reference value (deviation
measurement mode)
Functional description
The deviation measurement feature allows you to display deviation values
instead of actual measurements. A deviation is expressed as the difference
between the actual measurement and the stored reference value. The deviation
measurement feature is useful when you observe how a particular value of a
device/component changes under varying conditions of temperature,
frequency, bias, and other influences.
You can apply the deviation measurement feature to the primary or secondary
parameter or both. The deviation measurement feature supports the following
two modes:
ΔABS (absolute value) deviation measurement
Identifies and displays the difference between the actual measurement of
the DUT and the stored reference value. This value is calculated based on
the following formula:
Softkey Description
D.P. AUTO Automatically sets the decimal point
D.P. FIX Fixes the decimal point position
D.P. POS INCR+ Increments the position of decimal point
D.P. POS DECL- Decrements the position of decimal point
ΔABS = X-Y
X
Y
actual measurement of the DUT
stored reference value