Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 59
Setting up Measurement Conditions and Display
Setting Up Averaging Count
Setting Up Averaging Count
Functional description
The averaging feature of the E4981A allows you to obtain moving average
values of successive measurement results.
You can specify the averaging factor within the range of 1 to 256 in steps of 1.
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Use the cursor keys to select the AVG field
Step 3. Use the numeric keys or following softkeys:
Setting Up Trigger Delay Time
You can set up a wait time (trigger delay time) between when a trigger is
detected and when the measurement is started. You can set up the trigger
delay time within the range of 0 s to 1 s in steps of 100
μs.
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Use the cursor keys to select the TRIG DLY field.
Step 3. Use the numeric keys or following softkeys:
Softkey Description
ON Turns ON the averaging feature.
OFF Turns OFF the averaging feature.
INCR+ Increments the averaging factor in steps of 1.
DECR- Decrements the averaging factor in steps of 1.
Softkey Description
INCR++ Increments the trigger delay time in steps of 1 msec.
INCR+ Increments the trigger delay time in steps of 0.1 msec.
DECR- Decrements the trigger delay time in steps of 0.1 msec.
DECR-- Decrements the trigger delay time in steps of 1 msec.