EasyManua.ls Logo

Keysight E4981A - Electrical Characteristics; Output Signal

Keysight E4981A
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 161
Using Scanner Interface
Electrical Characteristics
Electrical Characteristics
Output signal
The output signals (/INDEX and /EOM) are available as open collector outputs
with photo coupler isolated. You can obtain each voltage output by connecting
a pull-up resistor (refer to Table 9-2) to the exterior of the E4981A.
Table 9-3 shows the electrical characteristics of the output signals. Figure 9-4
shows the circuit diagram of the output signals.
Table 9-2 Guide for pull-up resistor values
Pull-up voltage
[V]
Resistance value []
Typical resistance
Resistance
value []
Keysight part
number
5 1.7 k (5 V/3 mA) 1.78 k 0757-0278
9 3.0 k (9 V/3 mA) 3.16 k 0757-0279
12 4.0 k (12 V/3 mA) 4.22 k 0698-3154
15 5.0 k (15 V/3 mA) 5.11 k 0757-0438
Table 9-3 Electrical characteristics of the scanner interface output signals
Output signal
Output voltage [V] Maximum
current
[mA]
LOW HIGH
/INDEX, /EOM 0 - 0.5
EXT_DCV
1
6
1. EXT_DCV: +5 V - +15 V