Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 171
Specifications and Supplemental Information
Basic Specifications
Measurement display ranges
Table 10-1 shows the range of the measured value that can be displayed on
the screen.
Available measurement ranges
Table 10-2, Table 10-3 and Table 10-4 show recommended measurement
ranges (recommended for accurate measurement) and significant
measurement ranges (ranges that do not cause overload) for each
measurement value under the condition D (dissipation factor) ~ 0.5.
Table 10-1 Allowable measured value display range
Parameter Measurement display range
Cs, Cp ±1.000000 aF - 999.9999 EF
D ±0.000001 - 9.999999
Q ±0.01- 99999.99
Rs, Rp ±1.000000 aΩ - 999.9999 EΩ
G ±1.000000 aS - 999.9999 ES
D% ±0.0001%- 999.9999%
Table 10-2 Measurable capacitance ranges when measurement frequency is 120 Hz
Measurement value
Recommended
measurement range
Significant
measurement range
10 nF 0 nF - 15 nF 0 F - 15 nF
22 nF 15 nF - 33 nF 0 F - 33 nF
47 nF 33 nF - 68 nF 0 F - 68 nF
100 nF 68 nF - 150 nF 0 F - 150 nF
220 nF 150 nF - 330 nF 0 F - 330 nF
470 nF 330 nF - 680 nF 0 F - 680 nF
1 µF 680 nF - 1.5µF 0 F - 1.5 µF
2.2 µF 1.5 µF - 3.3 µF 0 F - 3.3 µF
4.7 µF 3.3 µF - 6.8 µF 0 F - 6.8 µF
10 µF 6.8 µF - 15 µF 0 F - 15 µF
22 µF 15 µF - 33 µF 0 F - 33 µF
47 µF 33 µF - 68 µF 0 F - 68 µF