126 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
7-
Figure 7-2 Example of limit range settings (case 2)
Setting up limit ranges
Notes on setup
• If you set up the upper limit value to a value equal to or less than the lower
limit value, the limit range is not used. This operation is equivalent to
setting the limit range to OFF.
• If BINs overlap, the resulting measurement is sorted into the BIN of the
smallest number (refer to Figure 7-5, “Sorting judgment flow,” on
page 133). Therefore, you need to set up the limit ranges starting from the
narrowest to the widest as shown in Figure 7-2.
• For the tolerance mode, the reference value is not required to be within the
limit range (between the lower limit value and upper limit value).
• Gaps between the limit ranges are allowed. Therefore, you can set up the
limit ranges as shown in Figure 7-3.
Table 7-2 Lower and upper limit values of limit ranges for Figure 7-2 (comparison of
modes)
Absolute mode
Tolerance mode (reference value: 1 nF)
Absolute Percent
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
BIN1 0.99 nF 1.01 nF –0.01 nF 0.01 nF –1% 1%
BIN2 0.98 nF 1.02 nF –0.02 nF 0.02 nF –2% 2%
BIN3 0.97 nF 1.03 nF –0.03 nF 0.03 nF –3% 3%