Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 127
Sorting Based on Measured Results (Comparator Function)
Setting Up Sorting Judgment Conditions
Figure 7-3 Example of gap between limit ranges
Procedure to set up BIN1 to BIN9 (ON/OFF and lower and upper limit
values)
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Use the cursor keys to select the BIN 1 field.
Step 4. Use the following softkeys:
Step 5. Repeat Step 3 to Step 4 using the cursor keys until you have turned ON the bin
9.
Step 6. Use the cursor keys to select the BIN 1 LOW field.
Table 7-3 Lower and upper limit values of limit ranges for Figure 7-3
Absolute mode
Tolerance mode (reference value: 1 nF)
Absolute Percent
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
Lower
limit
value
Upper
limit
value
BIN1 1 nF 1.02 nF 0 nF 0.02 nF 0% 2%
BIN2 0.97 nF 0.98 nF –0.03 nF –0.02 nF –3% –2%
BIN3 0.99 nF 1.03 nF –0.01 nF 0.03 nF –1% 3%
When the low limit value is greater than the upper limit, a warning
message “Improper high/low limits” is displayed.
Softkey Function
ON Turns ON the bin function.
OFF Turns OFF the bin function.