Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 67
Setting up Measurement Conditions and Display
Setting Up Condition To Make A Beep
Step 7. Use the cursor keys to select the CC1 TH2 field. TH2 is a parameter
proportional to the contact resistance of Lp or Lc.
Step 8. Use the softkeys or entry keys to enter the TH2 value according to Table 4-4.
When data is entered with the entry keys, the softkeys change to units labels
(m, x1).
Setting Up Condition To Make A Beep
The condition to make a beep sound differs depending on the beep mode, as
shown in Table 4-5.
Setup procedure
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Use the cursor keys to select the BEEP field.
Softkey Description
INCR++ Increments the threshold value as 0, 10 m, 20 m, 50 m, 100 m, 200 m,
500 m, 1.
INCR+ Increments the threshold value in steps of 10.
DECR- Decrements the threshold value in steps of 10.
DECR-- Decrements the threshold value 1, 500 m, 200 m, 100 m, 50 m, 20 m,
10 m, 0.
Table 4-5 Condition to make a beep
Mode Condition to make a beep
Off Never makes a beep.
Fail
When wrong key operation is
performed.
When an error, alarm, or other
message
is outputted.
When the sorting judgment result of
the comparator is OUT_OF_BIN,
AUX_BIN, No contact or OVLD.
Pass
When the sorting judgment result of
the comparator is from BIN1 to
BIN9.