52 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Selecting Measurement Parameters
4-
Figure 4-1 Relationship between equivalent circuit model and measurement parameters
For information on the selection criteria for the equivalent circuit model, refer
to “Selection criteria of parallel/series equivalent circuit models” on
page 247.
Setup procedure
For details on setup procedure, refer to “Setting up measurement parameters”
on page 40.
e4981aue0049
Cp
Cs Rs
Rp
D =
2Ǹf Cp Rp
1
=
2Ǹf Cp Rp
D =
2Ǹf Cs Rs
Q =
G =
Q =
1
D
Rp
1
1
D
2Ǹf Cs Rs
1
=
Parallel equivalent
circuit model
Series equivalent
circuit model
f: M easurem en t s ignal frequency ( 120 Hz / 1 kH z / 1 M Hz )