Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 173
Specifications and Supplemental Information
Basic Specifications
Measurement accuracy
The measurement accuracy is defined when all of the following conditions are
met.
• Warm-up time: 30 minutes or longer
• Ambient temperature: 18°C - 28°C
• Execution of OPEN correction
• Execution of Cable correction for 1 MHz measurement
• Measurement cable length: 0 m, 1 m, or 2 m (16048A/B/D)
1
• D (dissipation factor) ~ 0.5
Table 10-4 Measurable capacitance ranges when measurement frequency is 1 MHz,
1 MHz ± 1%, 1 MHz ± 2%
Measurement value
Recommended
measurement range
Significant
measurement range
1 pF 0 F - 1.5 pF 0 F - 1.5 pF
2.2 pF 1.5 pF - 3.3 pF 0 F - 3.3 pF
4.7 pF 3.3 pF - 6.8 pF 0 F - 6.8 pF
10 pF 6.8 pF - 15 pF 0 F - 15 pF
22 pF 15 pF - 33 pF 0 F - 33 pF
47 pF 33 pF - 68 pF 0 F - 68 pF
100 pF 68 pF - 150 pF 0 F - 150 pF
220 pF 150 pF - 330 pF 0 F - 330 pF
470 pF 330 pF - 680 pF 0 F - 680 pF
1 nF 680 pF - 1.5 nF 0 F - 1.5 nF
1. The outer conductor resistance of cable requires the following conditions:
16048A/B: 62 mΩ or below
16048D: 90 mΩ or below